This machine is mainly suitable for shock resistance tests in electronic products, aerospace, shipbuilding, military industry, auto parts, transportation and other fields. By choosing different waveform generators, impact tests such as half sine wave, trailing peak sawtooth wave or square wave can be carried out.
Environmental requirements:
1. Site temperature: 5℃~+28℃ (average temperature within 24 hours≤28℃);
2. Relative humidity: ≤85%RH;
3. Air source: 0.5--0.8MPa
4. Power supply conditions: single-phase three-wire;
5. Voltage range: AC220V;
6. Allowable frequency fluctuation range: (50±0.5) Hz;
7. The grounding resistance of the protective earth wire is less than 4Ω;
8. The user is required to configure an air or power switch with corresponding capacity for the equipment at the installation site, and this switch must be independent for the use of this equipment.
Technical parameter:
Work platform |
250 × 250mm |
350 × 350 mm |
Maximum load |
10kg |
30kg |
Work surface material |
High-strength imported high-quality |
|
Aluminum alloy |
Maximum drop height 1500mm |
|
Peak acceleration |
20-1000 G |
|
Pulse duration |
0.5-20ms |
|
Equipment size |
650 × 700 × 2370mm |
|
Total equipment |
820 kg |
|
Aeasuring system |
||
Input channel |
2 channels |
|
Sampling frequency |
192KHz |
|
Input form |
voltage, charge |
|
Voltage range |
-10V--+10V |
|
Coupling mode |
AC, DC, ICP |
|
Pulse duration |
50—1ms |
|
Shock waveform |
half sine |
|
Communication interface |
USB2.0 |
|
Support standards |
National standard, national military standard, U.S. military standard, user-defined standard |
|
Operating system |
Microsoft Windows7/10 |
|
System acceleration sensor |
||
Brand |
LAB |
|
Output method |
Charge type |
|
Sensitivity |
3.93pC/g |
|
Frequency range |
0.5-12KHz |
|
Acceleration range |
±2500G |
|
Working environment |
-40~+160℃ |
|
Other instructions |
||
Power supply |
AC 3-phase 380V±10% 5KVA |
|
Compressed air |
0.5-0.8Mpa |
|
Temperature |
RT~40℃ |
|
Humidity |
25℃<85RH% |